71
CVM
k
2
CONFIGURATION
4.7.6.4. Static outputs conguration
The outputs of the expansion card are congurating in that screen. The transistor are called
OUT 01, OUT 02, OUT 03 and OUT 04.
On this screen you congure alarms equations that are applied to select the outputs of the
device. You can set up equations with functions AND (*) and / or OR (+) in one or more of the
16 pre-congured alarms (see section 4.7.6.3 Alarm conguration), to activate each of the 4
output transistor of the team.
To modify the conguration parameters of the card, you must press the EDIT (F4). Select the
output you want to congure and press SET to enter editing.
1
A. OUTPUTS T. OUTPUTS
2345678 9 1 2345678 9
1- COMMON
2- Analogic output 1
3- COMMON
4-
Analogic output 2
5- COMMON
6-
Analogic output 3
7- COMMON
8-
Analogic output 4
9- No used
1- COMMON
2- COMMON
3-
4-
5-
6-
7-
8-
9- Transistor ouput 4
COMMON
COMMON
COMMON
Transistor ouput 1
Transistor ouput 2
Transistor ouput 3
4.7.6.5. Outputs wiring
The layout of the outputs of the expansion card is as follows
Commenti su questo manuale