Press key to access the next programming step.
4�9�29� Locking the programming
This screen is for protecting the data configured in the
programming menu.
Press key
for 3 seconds to edit the locking/unlocking
selection. The prog icon will be displayed on the bottom of the
screen.
Press key
to browse the two options:
When you enter the programming menu you can view and modify the programming. Icon
on the display indicates the permanently locked status.
When you enter the programming you can view the programming but not modify it. Icon
indicates the locking status. Enter the password to modify the programming values.
To validate the data, press for 3 seconds and the prog icon will disappear from the display.
Press the key to enter the password for locking and unlocking the programming:
On this screen you enter the password for locking and unlocking
the programming.
Press key
for 3 seconds to edit the password selection. The
prog icon will be displayed on the bottom of the screen.
To enter or modify the value, press the
key repeatedly,
increasing the value of the ashing digit.
When the desired value is shown on the screen, move onto the next digit by pressing the key
to modify the other values.
If you press the key after changing the last digit, it will jump back to the rst digit so you
can modify the previously programmed values again.
To validate the data, press for 3 seconds and the prog icon will disappear from the display.
Default password: 1234.
This value may only be modified through communications. See “4.10.3.6.16. Password
conguration.”
Press the key to exit the setup menu.
Note: The unit will restart after exiting the setup menu if the RS-485 communications
parameters have been modied.