Circutor CVM-C10 Series Manuale Utente Pagina 41

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 82
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 40
Single-phase network measurement, phase to neutral, with a 2-wire connection.
To validate the data, press for 3 seconds and the prog icon will disappear from the display.
Press key to access the next programming step.
4�9�9� Maximum demand integration period
The maximum demand integration period is programmed in
minutes on this screen.
Press key
for 3 seconds to edit the integration period value.
The prog icon will be displayed on the bottom of the screen.
To enter or modify the value, press the
key repeatedly,
increasing the value of the ashing digit.
When the desired value is shown on the screen, move onto the next digit by pressing the key
to modify the other values.
If you press the key after changing the last digit, it will jump back to the rst digit so you
can modify the previously programmed values again.
To validate the data, press for 3 seconds and the prog icon will disappear from the display.
The programmed value will be deleted if the entered value is higher than the maximum
programming value.
Maximum programming value: 60.
Minimum programming value: 1.
Press key to access the next programming step.
4�9�10� Deleting maximum demand
On this screen you select whether or not to delete the maximum
demand.
Press key
for 3 seconds to edit the deletion selection. The
prog icon will be displayed on the bottom of the screen.
Press key
to browse the two deletion options: Yes or No.
To validate the data, press for 3 seconds and the prog icon will disappear from the display.
Press key to access the next programming step
41
Instruction Manual
CVM-C10
Vedere la pagina 40
1 2 ... 36 37 38 39 40 41 42 43 44 45 46 ... 81 82

Commenti su questo manuale

Nessun commento